BIEN AIR BLACK PEARL ECO
Price: 6500.0 EGP
SIMPLICITY ITSELF Experience Bien-Air’s Swiss-made quality at a value defying all logic. Boralina and Blackpearl Eco let you push the limits of your objectives with more potential and ability than any high-speed handpiece in their class. The opportune support you can depend on when you need...

SIMPLICITY ITSELF

Experience Bien-Air’s Swiss-made quality at a value defying all logic. Boralina and Blackpearl Eco let you push the limits of your objectives with more potential and ability than any high-speed handpiece in their class. The opportune support you can depend on when you need to be ready for anything.

Accu-Spray Sep.1 system

This ultra-precise nebulization system is composed of one separate air and water spray providing adequate cooling to the bur and treatment site.


Accu-Chuck™
bur-locking and rotation drive mechanism

This proven bur-locking and vibration-canceling system vouches for optimal stability and precision, no matter the rotational speed. Assembled with Bien-Air's Soft Push™ button, bur release is quick and easy.

Soft Push

Bien-Air’s unique push-button insert locking mechanism. Light pressure is all that is needed. Bur changing is quick and comfortable. This mechanism guarantees firm yet precise locking.


TECHNICAL DATA

Black Pearl Eco 2-3 wayBlack Pearl ECO 4 way
Power (W)1313
Spray outlet2 x 1 separate2 x 1 separate
Noise level (dBa)6262
Head diameter (mm)1212
Head size (mm)13.8 (22.5 mm with bur 19 mm)13.8 (22.5 mm with bur 19 mm)
Rotation speed (rpm)310’000310’000
Light--
Light intensity (klux)--
Ceramic ball bearingsNoNo
Weight (g)5050
Coupling2-3 way4 way
Warranty1 to 2 years11 to 2 years1
ReferenceBlack Pearl Eco 2-3 way 1600039-001Black Pearl ECO 4 way 1600040-001

MAINTENANCE

MAINTENANCE OF TURBINES.

PREPARATION
Preparation
A.

REMOVING THE INSTRUMENT

  • Remove the rotary instrument from the chuck mechanism
Tissue_rubbing
B.

PRE-DISINFECTION

  • External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING
Cleaning_A
A.

CLEAN UNDER RUNNING WATER AT <38°C

Cleaning_B
B.

REMOVING DIRT/DEPOSITS

  • Carefully remove all traces of dirt or deposits from the spray nozzles using the Bien-Air cleaning wire
  • Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

Cleaning_CSpraynet-Tissue-spray_CTissue_rubbing_C
C.

INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

  • Shake the Spraynet before use
  • Use the Spraynet attachment 1
    (for 2-/3-way and 4-way turbines, see inset below)
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
  • Clean the exterior of the turbine with a cloth soaked in Spraynet
FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
Cleaning_C-primeBlack-Pearl-Eco_2-3waysBlack-Pearl-Eco_4-ways
C'.

CLEANING THE INSIDE WITH SPRAYNET FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

  • Use the Spraynet attachment 2
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second

Spray into the opening as shown in the illustrations for “2-/3-way” and “4-way”

DISINFECTION
Disinfectant-Tissue-spray_ATissue_rubbing_A
A.

DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer's instructions for use)

Do not immerse the instrument in a disinfectant bath

OR
Thermal-disinfector
B.

THERMAL DISINFECTION (follow the manufacturer's instructions for use)

LUBRICATION
Lubrification_A1Lubrification_A2
A.

LUBRICATION WITH LUBRIMED

OR
Lubrification_B1Support_B
B.

LUBRICATION WITH LUBRIFLUID

  • Shake the Lubrifluid before use
  • Use the Lubrifluid attachment 2, 3 ou 4depending on the type of instrument * (For 2-/3-way and 4-way turbines, see inset below)
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
  • Then, run an operating test for 30 seconds at low speed to remove any surplus lubricant
  • Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out
FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
LubrificationBlack-Pearl-Eco_2-3waysBlack-Pearl-Eco_4-ways
B'.

INTERNAL LUBRICATION WITH LUBRIFLUID FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

  • Use the Lubrifluid attachment 2
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second

Spray into the opening as shown in the illustrations for “2-/3-way” and “4-way”

STERILISATION
Sterilization
A.

PLACE THE TURBINE IN A STERILISATION BAG

Sterilizer
B.

STERILISE IN ACCORDANCE WITH THE MANUFACTURER'S INSTRUCTIONS

  • Instruments bearing the "135°C" symbol can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated