Trident X-Ray RiX-70 DC Mobile
Price: 46000.0 EGP
VoltageAC 100-240 ±
Frequency50/60 Hz
Electric current2.1 A(@240V), 5.5 A (@100V)
Max power consumption640 VA
Fuse6.3 AT
Connection7 wires cable with double insolation

Automatic Voltage Transformer


RiX-70 DC appropriately works in a range of 100 to 240 V. The electrical system automatically detects the power wattage, allowing a safe and proper functioning with every voltage.

0.4 mm Focal Spot

The base for an accurate diagnostic is a good image. RiX-70 DC with focal spot 0,4 mm gives you incredibly clear images with a high contrast range.


Reduction of radiation dose

The high frequency generator allows radiation stability eliminating the soft radiation that does not produce images but can be harmful for the patient.


Tube Voltage Selector

RiX-70 DC operates either 60, 65 or 70 kV allowing to reach excellent results to satisfy your needs:

Voltage 60 kV for caries detection or periodontal applications; 70 kV for endodontic applications, including visualization of canals, detection of root fractures and diagnosis of periapical lesions due to pulpal inflammation.

Small details, great results.

Automatic setting of exposure time from 0.01 to 2.0 seconds, depending on the patient and tooth characteristics


Easy-to-read main control panel, with quickly identifiable icons.

From the control panel you can easily:

  • Select the exposure parameters according to the patient morphology, tooth type and image receptor.
  • Adapt the radiation dose to lower or higher values.
  • Modify the exposure time for digital or film.
  • Select from 60, 65 or 70 kVp

Modern and clean design to offer you:
  • Compact and lightweight tube head.
  • Steady and strong support arm.
  • Hand switch provided with 3 mt coiled cord.
  • Clear and intuitive remote control (optional)
  • Rectangular collimator 35 x 43 mm (optional).
  • Additional keypad installed near the tube head, for manual adjustment of the exposure time and tooth type.

Monoblock

Tube Voltage60 kVp ± 8 %
65 kVp ± 8 %
70 kVp ± 8 %
Anodic current7 mA± 10 %
Output radiation linearity< 0.2 secondo IEC 60601-2-63
Duty cycle1/30 6 seconds minimum
Nominal power490 W (70 kVp - 7 mA)
Total filtration>1.5 mm Al eq.
Current transformer insulationOil bath
CoolingAs per convention
Focal spot0.4 mm (IEC 60336)
Leakage radiation<0.15 mGy/h at 1 m
Max thermal capacity of monoblock130 kJ
Focus to skin distance20 cm (8 5/8 inch)
Output radiation FieldΦ 60mm (2 3/8 inch)
Bean limiting device (additional)Rectangular 36 x45 mm
Dose intensity accuracy< 50%


X-Ray Tube

ModelToshiba D-045CEI OX70-4
Inherent filtration< 1 mm Al eq0.5 mm Al eq.
Target angle16°20°
Filament currentMax. 3 A1.5 – 2.1 A
Max filament current3 V1.7 – 3.1 V
Anodic thermal capacity4.3 kJ7 kJ
Max anodic dissipation100 W110 W
X RAY beam covering at 200 mm from the focus>= 70 mm>= 70 mm
Focal spot0.4 mm (IEC 60336)0.4 mm (IEC 60336)
Target materialTungstenTungsten
Max nominal voltage70 kV70 kV